Description and commissioning of NEXT-MM prototype: first results from operation in a Xenon-Trimethylamine gas mixture
dc.contributor.author
dc.date.accessioned
2019-09-18T06:30:11Z
dc.date.available
2019-09-18T06:30:11Z
dc.date.issued
2014-03-18
dc.identifier.issn
1748-0221
dc.identifier.uri
dc.description.abstract
A technical description of NEXT-MM and its commissioning and first performance is reported. Having an active volume of ~ 35 cm drift × 28 cm diameter, it constitutes the largest Micromegas-read TPC operated in Xenon ever constructed, made by a sectorial arrangement of the 4 largest single wafers manufactured with the Microbulk technique to date. It is equipped with a suitably pixelized readout and with a sufficiently large sensitive volume ( ~ 23 l) so as to contain long ( ~ 20 cm) electron tracks. First results obtained at 1 bar for Xenon and Trymethylamine (Xe-(2%)TMA) mixture are presented. The TPC can accurately reconstruct extended background tracks. An encouraging full-width half-maximum of 11.6 % was obtained for ~ 29 keV gammas without resorting to any data post-processing
dc.description.sponsorship
The NEXT Collaboration acknowledges funding support from the following agencies and institutions:
the Spanish Ministerio de Economía y Competitividad under grants CONSOLIDERIngenio
2010 CSD2008-0037 (CUP), Consolider-Ingenio 2010 CSD2007- 00042 (CPAN), and
under contracts ref. FPA2008-03456, FPA2009-13697-C04-04; FCT(Lisbon) and FEDER under
grant PTDC/FIS/103860/2008; the European Commission under the European Research Council
T-REX Starting Grant ref. ERC-2009-StG-240054 of the IDEAS program of the 7th EU Framework Program; Director, Office of Science, Office of Basic Energy Sciences, of the US Department
of Energy under contract no. DE-AC02-05CH11231. Part of these grants are funded by the European
Regional Development Fund (ERDF/FEDER). J. Renner (LBNL) acknowledges the support
of a US DOE NNSA Stewardship Science Graduate Fellowship under contract no. DE-FC52-08NA28752
dc.format.mimetype
application/pdf
dc.language.iso
eng
dc.publisher
Institute of Physics (IOP)
dc.relation.isformatof
Versió preprint del document publicat a: https://doi.org/10.1088/1748-0221/9/03/P03010
dc.relation.ispartof
© Journal of Instrumentation, 2014, vol. 9, p.P03010
dc.relation.ispartofseries
Articles publicats (D-EMCI)
dc.rights
Tots els drets reservats
dc.subject
dc.title
Description and commissioning of NEXT-MM prototype: first results from operation in a Xenon-Trimethylamine gas mixture
dc.type
info:eu-repo/semantics/article
dc.rights.accessRights
info:eu-repo/semantics/openAccess
dc.type.version
info:eu-repo/semantics/submittedVersion
dc.identifier.doi
dc.identifier.idgrec
021450