Scatter reduction for grid-less mammography using the convolution-based image post-processing technique
dc.contributor.author
dc.date.accessioned
2017-10-24T10:26:12Z
dc.date.available
2017-10-24T10:26:12Z
dc.date.issued
2017-01-01
dc.identifier.issn
1605-7422
dc.identifier.uri
dc.description.abstract
X-ray Mammography examinations are highly affected by scattered radiation, as it degrades the quality of the image and complicates the diagnosis process. Anti-scatter grids are currently used in planar mammography examinations as the standard physical scattering reduction technique. This method has been found to be inefficient, as it increases the dose delivered to the patient, does not remove all the scattered radiation and increases the price of the equipment. Alternative scattering reduction methods, based on post-processing algorithms, are being investigated to substitute anti-scatter grids. Methods such as the convolution-based scatter estimation have lately become attractive as they are quicker and more flexible than pure Monte Carlo (MC) simulations. In this study we make use of this specific method, which is based on the premise that the scatter in the system is spatially diffuse, thus it can be approximated by a two-dimensional low-pass convolution filter of the primary image. This algorithm uses the narrow pencil beam method to obtain the scatter kernel used to convolve an image, acquired without anti-scatter grid. The results obtained show an image quality comparable, in the worst case, to the grid image, in terms of uniformity and contrast to noise ratio. Further improvement is expected when using clinically-representative phantoms
dc.description.sponsorship
Oliver Díaz is supported by the European Union within the Marie Sklodowska-Curie Innovative Training Networks (H2020-MSCA-IF-2014 SCARtool project, reference 657875) and the Ministry of Economy and Competitiveness of Spain, under project reference DPI2015-68442-R
dc.format.mimetype
application/pdf
dc.language.iso
eng
dc.publisher
SPIE Proceedings
dc.relation
info:eu-repo/grantAgreement/MINECO//DPI2015-68442-R/ES/ANALISIS DE IMAGENES INTELIGENTE PARA LOS RETOS EN EL CRIBADO DE CANCER DE MAMA/
dc.relation.isformatof
Versió postprint del document publicat a: https://doi.org/10.1117/12.2255558
dc.relation.ispartof
© Flohr, Thomas G.; Lo, Joseph Y.; Schmidt, Taly Gilat (Eds.). (2017). SPIE Medical Imaging 2017: Physics of Medical Imaging: 11-16 February 2017, Orlando, FLA, USA. (SPIE Proceedings: , vol. 10132)
dc.relation.ispartofseries
Articles publicats (D-ATC)
dc.rights
Tots els drets reservats
dc.subject
dc.title
Scatter reduction for grid-less mammography using the convolution-based image post-processing technique
dc.type
info:eu-repo/semantics/article
dc.rights.accessRights
info:eu-repo/semantics/openAccess
dc.embargo.terms
Cap
dc.relation.projectID
info:eu-repo/grantAgreement/EC/H2020/657875/EU/Scattered radiation reduction tool to improve computer-aided diagnosis performance in digital breast tomosynthesis/SCARtool