Cr speciation in water samples by dispersive liquid-liquid microextraction combined with total reflection X-ray fluorescence spectrometry
Texto Completo
Cr-speciation-in-water.pdf
Solicita copia
Al rellenar este formulario estáis solicitando una copia del artículo, depositado en el repositorio institucional (DUGiDocs), a su autor o al autor principal del artículo. Será el mismo autor quien decidirá enviar una copia del documento a quien lo solicite si lo considera oportuno. En todo caso, la Biblioteca de la UdG no interviene en este proceso ya que no está autorizada a facilitar artículos cuando éstos son de acceso restringido.
Compartir
A novel method based on dispersive liquid-liquid microextraction (DLLME) and total-reflection X-ray fluorescence spectrometry (TXRF) is proposed for the determination and preconcentration of inorganic Cr species (Cr(VI) and Cr(III)) in water samples. Parameters affecting the extraction procedure for Cr speciation (pH, sample volume, disperser solvent, complexing agent, organic solvent, temperature) and TXRF conditions (sample volume, drying mode, measurement time, internal standardization) have been carefully evaluated to ensure the highest sensitivity for Cr determination. It was found that the minimum Cr content that can be detected was 0.8 μg L- 1. This value is significantly lower than the maximum Cr content permissible in drinking waters and it is better or comparable with those obtained in other published works based on the use of solid phase extraction or liquid phase microextraction in combination with atomic absorption spectrometry techniques. A good linearity (R2 = 0.9937) was obtained in the range of 5 to 4000 μg L- 1 demonstrating the suitability of the DLLME + TXRF for both the analysis of drinking and waste water samples. Additional advantages of the DLLME-TXRF systems include the use of very small amount of reagents and the simplicity and low cost of operation of the benchtop instrument used (no cooling media and gas consumption are needed)
Tots els drets reservats