Real-Time Continuous Glucose Monitoring in an Intensive Care Unit: Better Accuracy in Patients with Septic Shock

Compartir
This study assessed the accuracy of real-time continuous glucose monitoring system (RTCGMS) devices in an intensive care unit (ICU) to determine whether the septic status of the patient has any influence on the accuracy of the RTCGMS. Subjects and methods: In total, 41 patients on insulin therapy were included. Patients were monitored for 72 h using RTCGMS. Arterial blood glucose (ABG) samples were obtained following the protocol established in the ICU. The results were evaluated using paired values (excluding those used for calibration) with the performance assessed using numerical accuracy. Nonparametric tests were used to determine statistically significant differences in accuracy. Results: In total, 956 ABG/RTCGMS pairs were analyzed. The overall median relative absolute difference (RAD) was 13.5%, and the International Organization for Standardization (ISO) criteria were 68.1%. The median RADs reported for patients with septic shock, with sepsis, and without sepsis were 11.2%, 14.3%, and 16.3%, respectively (P<0.05). Measurements meeting the ISO criteria were 74.5%, 65.6%, and 63.7% for patients with septic shock, with sepsis, and without sepsis, respectively (P<0.05). Conclusions: The results showed that the septic status of patients influenced the accuracy of the RTCGMS in the ICU. Accuracy was significantly better in patients with septic shock in comparison with the other patient cohorts ​