When filling the form you are requesting a copy of the article, that is deposited in the institutional repository (DUGiDocs), at the autor or main autor of the article. It will be the same author who decides to give a copy of the document to the person who requests it, if it considers it appropriate. In any case, the UdG Library doesn’t take part in this process because it is not authorized to provide restricted articles.
Share
In the present work, the analytical performance of a methodology based on a preconcentration using activated thin layers followed by high-energy polarised beam energy-dispersive XRF spectrometry (HE-P-EDXRF) for the determination of trace amounts of Cd in saline solutions is presented. Linear calibration curves were obtained over the range 5-8000 μg L-1 of Cd. When comparing the developed method to the direct analysis of Cd in sea water, the detection limits were improved by several orders of magnitude, with values of 2.25 μg L-1 (Al2O3 target) and 0.075 μg L -1 (CsI target), respectively. The accuracy was checked by analyzing spiked natural sea water at Cd concentrations of 20, 100 and 600 μg L -1 and results were compared to those obtained with anodic stripping voltametry and no significant differences at the 95% confidence level were found. The method has a good reproducibility over the investigated period of time and the precision of the method was better than 10% ​